The Dulas VC225ILR is the largest Hot Zone rated large capacity next generation ‘icelined’ vaccine refrigerator. A bespoke design for use in the cold chain – the result of over 30 years’ experience gained in supplying refrigerators to the most remote corners of the earth.

Bringing Freeze-free technology to mains powered refrigerators, the water ‘ice lining’ in the VC225ILR is replaced with our advanced non-corrosive phase change material (PCM) which becomes an ILR with no vaccine freeze risk.

At 184.5 litres vaccine capacity the VC225ILR is ideal for larger health facilities and is supplied with a built in power regulator and voltage stabiliser.

Features and Benefits
Technical Specification
Downloads
  • 184.5 litre vaccine storage capacity
  • +5°C to +43°C extended operating temperature range
  • Works with only 6 hours of electricity each day
  • 94 hours holdover at +43°C
  • Corrosion tested to DIN 8985
  • In-built handle with lock
  • In-built solar powered temperature display
  • All-metal, high strength hinges
  • Heavy duty castors to make handling easier
  • User-independent Grade A freeze protection
  • Environmentally friendly: R600a refrigerant with low global warming potential and innovative insulation technology, giving exceptional energy efficiency
  • Packaging tested to industry standards
  • Tested to survive ambient storage temperatures up to +70°C
  • British made
  • CFC-free
Performance  
Net vaccine storage capacity: 184.5 litre
Climate zone: Hot Zone (+43⁰C)
Minimum rated ambient temperature: +5⁰C
Holdover time: 94 hours at +43°C
Hours of electricity required to operate: > 6 hours per 24hrs
Recommended solar array size: > 400Wp
Dimensions  
External: 986 x 1287 x 780 (mm)
Shipping weight: 129 kg
Gross volume: 203 litre
Technical Details  
Nominal Voltage: 230V
Operating voltage Range: 24-45Vdc
Nominal frequency: 50Hz
Operating frequency range 45 to 65Hz
Compressor: Danfoss
Refrigerant: R600a
WHO Pre-qualification  
Test procedure: E003/RF05.3-VP.3
Performance specification: E003/RF05.3
PQS reference: E003/072