VC88SDD Solar Direct Drive Refrigerator

VC88SDD Solar Direct Drive Refrigerator

The VC88SDD is an advanced mid-sized WHO accredited 88 litre capacity Solar Direct Drive (SDD) Vaccine Refrigerator.

Offering class leading total cost of ownership and cost per litre of vaccine storage it is the perfect choice for upgrading your cold chain.

Utilising our optimised solar technology the VC88SDD offers the perfect balanced of array size and autonomy to deliver exceptional reliability whatever the weather.

Fully GAVI CCEOP compliant with the highest level of Grade A Freeze Protection and a +5C to +43C extended temperature operating range.

Features and Benefits
Technical Specification
Downloads
  • 88 litre vaccine storage capacity
  • +5°C to +43°C extended operating temperature range
  • Works at WHO solar reference period 3.5kWh/m2 /24h
  • 90hr 02min holdover at +43°C
  • 78hr 09min autonomy at +43°C
  • Corrosion tested to DIN 8985
  • In-built handle with lock
  • In-built solar powered temperature display
  • All-metal, high strength hinges
  • Heavy duty castors to make handling easier
  • User-independent Grade A freeze protection
  • Completely battery free, with plug and play connectors for easy installation
  • Environmentally friendly: R600a refrigerant with low global warming potential and innovative insulation technology, giving exceptional energy efficiency
  • Packaging tested to industry standards
  • Tested to survive ambient storage temperatures up to +70°C
  • Stacking baskets for easy-access stock management
  • British made
  • CFC Free
Performance  
Net vaccine storage capacity: 88 litre
Climate zone: Hot Zone (+43⁰C)
Minimum rated ambient temperature: +5⁰C
Holdover time: 90hr 02min
Autonomy time: 78hr 09min
Recommended solar array size: > 400Wp
Dimensions  
External: 959 x 1282 x 740 (mm)
Shipping weight: 129 kg
Gross volume: 205 litre Refrigerator
Storage Baskets 4
Technical Details  
Operating voltage: 24-45Vdc
Minimum starting voltage: 24Vdc
Minimum starting power 50w refrigerator
Minimum solar radiation for continuous running: 125 W/m2 (using 400 Wp array)
Compressor: Danfoss / Secop BD35K
Refrigerant: R600a
WHO Pre-qualification  
Test procedure: E003/RF05.3-VP.3
Performance specification: E003/RF05.3
PQS reference: E003/059

Download the VC88SDD Datasheet